Specifications | BiPAP Vision®

Modes

Mode

Control Range

Control Increments

CPAP 4 to 20 cmH2O 1 cmH2O
S/T 4 to 40 cmH2O 1 cmH2O
PAV /T Quick Start Menu Obstructive, Restrictive, Mixed, Normal

Top of Page

Control Ranges & Increments: Parameters

Parameters

Control Range

Control Increment

IPAP 4 to 40 cmH2O 1 cmH2O
EPAP 4 to 20 cmH2O 1 cmH2O
CPAP 4 to 20 cmH2O 1 cmH2O
Rate 4 to 40 BPM 1 BPM
Timed Inspiration 0.5 to 3.0 sec 0.1 sec
IPAP >Rise Time 0.05 to 0.40 sec 4 set points: .05, .1, .2, .4 sec
Oxygen Concentration 21 to 100% 4% from 21 to 25%
5% from 25 to 100%

Top of Page

Control Ranges & Increments: Alarms

Alarm

Control Range

Control Increment

High Pressure 5 to 50 cmH2O 1 cmH2O
Low Pressure Disabled;0 to 40 cmH2O 1 cmH2O
Low Pressure Delay 0 to 60 sec 1 sec
Apnea Disabled; 20 to 40 sec 4 set points:
Disabled, 20, 30, 40 sec
Low Minute Ventilation Disabled; 1 to 99 L/min 1 L/min
High Rate 4 to 120 BPM 1 BPM
Low Rate 4 to 120 BPM 1 BPM

Top of Page

Display Ranges & Increments

Parameters

Control Range

Control Increment

IPAP 0 to 50 cmH2O 1 cmH2O
EPAP 0 to 50 cmH2O 1 cmH2O
CPAP 0 to 50 cmH2O 1 cmH2O
Rate 0 to 150 BPM 1 BPM
Exhaled Tidal Volume (VT) 0 to 4000 ml 1 ml
Minute Ventilation (Min Vent) 0 to 99 L/min 1 L/min
Total Leak (Tot Leak) 0 to 300 L/min 1 L/min
Patient Leak(Pt. Leak) 0 to 300 L/min 1 L/min
Peak Inspiratory Pressure (PIP) 0 to 50 cmH2O 1 cmH2O
Percent of Patient Triggered Breaths (Pt. Trig) 0 to 100% 1%
Ti/Ttot 0 to 100% 1%